This document specifies the reference time input of a RUB GT module. If the reference input signals are within the limits of the following "general specifications" of the RUB GT module, it can operate in the reference time mode without constraints. 1. Serial data receive
Electrical specifications:
2. PPS input Input a positive or negative going pulse each second. Electrical specifications:
3. Time, date and status string of the serial data interface The data string shall be sent periodically (every second). The HH:MM:SS time information shall be in an up-counting sequence without time jumps. The complete data string shall fit between two timing reference signals (see below). RUB GT can receive the following data strings:
This list will be updated when new data strings are added. 4. Timing reference RUB GT needs a timing reference. This can be the PPS signal or - if no PPS signal is available - the start word of the serial data string. The time between the two timing reference signals shall be one second and shall not vary more than 1%. 5. List of recommended devices
This list is to be continued. 6. Application diagrams Reference time input with PPS and serial data. Stand-alone unit.
Reference time input with PPS and serial data. Unit receives power from the GT module.
|
||||||||||||||||||||
![]() |
|